3.2 Remote control commands

784 KFP Titrino

89

.Feed:

Length of the time axis. Depending on the printer, the measure in
cm may not always be correct.

Curve length

0.01

100 cm

0.1

10 cm

0.5

2 cm

1

1 cm



3.2.2.94. Setup.PowerOn

$G

Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.


3.2.2.95. Setup.Initialise

$G

Setup.Initialise.Select

ActMeth, Silo, Config,

Assembly, Setup, All

Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for

the data output, operands C01...C19.

Silo:

The silo memory is deleted. Same function as delete entire silo.

Config:

All values under &Config.

Assembly: All values under &Assembly.
Setup:

All values under &Setup.

All:

Values of the entire tree (except silo and method memory).

The action must be triggered with &Setup.Initalise $G.


3.2.2.96. Setup.RamInit

$G

Initializes instrument, see page 110. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted. The user memory contains the default user methods from Metrohm.


3.2.2.97. Setup.InstrNo

$G

Setup.InstrNo.Value

serial number, 8 ASCII characters

Instrument identification for report output.
Set the value with &Setup.InstrNo $G .


3.2.2.98. Diagnose.Report

$G

Output of the report containing the adjustment parameters. The Titrino has to
be in its inactive basic state.