3.2 Remote control commands
784 KFP Titrino
89
.Feed:
Length of the time axis. Depending on the printer, the measure in
cm may not always be correct.
Curve length
0.01
100 cm
0.1
10 cm
0.5
2 cm
1
1 cm
3.2.2.94. Setup.PowerOn
$G
Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.
3.2.2.95. Setup.Initialise
$G
Setup.Initialise.Select
ActMeth, Silo, Config,
Assembly, Setup, All
Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Silo:
The silo memory is deleted. Same function as delete entire silo.
Config:
All values under &Config.
Assembly: All values under &Assembly.
Setup:
All values under &Setup.
All:
Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
3.2.2.96. Setup.RamInit
$G
Initializes instrument, see page 110. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted. The user memory contains the default user methods from Metrohm.
3.2.2.97. Setup.InstrNo
$G
Setup.InstrNo.Value
serial number, 8 ASCII characters
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
3.2.2.98. Diagnose.Report
$G
Output of the report containing the adjustment parameters. The Titrino has to
be in its inactive basic state.