beautypg.com

2 description of the remote control commands – Metrohm 784 KFP Titrino User Manual

Page 72

background image

3.2 Remote control commands

784 KFP Titrino

67

3.2.2 Description of the remote control commands

3.2.2.1.

Mode

$G, $S, $H, $C

Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and
continue with $C.
$G also serves to continue after inquiries of identifications and sample size
after the start (see 3.2.2.26).


3.2.2.2.

Mode.QuickMeas

$G, $S

Start and stop of a measurement in the basic mode with the parameters
(measured quantity, measuring input) of the current method. Corresponds to
the key.
With an ongoing measurement, the current mode can be started. This stops
the measurement automatically.


3.2.2.3.

Mode.Select

KFT

Mode.KFTQuantity

Ipol, Upol

Selection of the standard mode. Mode and

If a method is selected from the method memory, the nodes &Mode.Select
and &Mode.XXXQuantity are overwritten with mode and measured quantity of
the corresponding user method.

the measured quantity belong to

the complete selection.



3.2.2.4.

Mode.Name

read only

Name of the current method in the working memory. $Q sends 8 ASCII cha-
racters. Standard methods carry the name ********. The node can be set
read + write, see 3.2.2.62.


3.2.2.5.

Mode.Def.Formulas.1.Formula

EPX, CXX, RSX, +, –, ∗, /, (, )

Mode.Def.Formulas.1.TextRS

up to 8 ASCII characters

Mode.Def.Formulas.1.Decimal

0...2...5

Mode.Def.Formulas.1.Unit

up to 6 ASCII characters

Mode.Def.Formulas.1.Limits

ON, OFF

Mode.Def.Formulas.1.LoLim

0...±999 999

Mode.Def.Formulas.1.UpLim

0...±999 999

Mode.Def.Formulas.1.Output

active, pulse, OFF

Mode.Def.Formulas.2.Formula

etc. up to .9

Entry of formulas. Rules for formula entry, see page 25.
Example: "(EP2–EP1)∗C01/C00"
In addition to the formula, a text for result output, the number of decimal plac-
es and a unit for the result output can be selected. "No unit" is selected with
the blank string.
In place of "RSX", a result name may be entered (.TextRS). This name is out-
putted in the report full, short, scalc full and scalc srt. It is used for the result
and the corresponding mean value.