3.2 Remote control commands

784 KFP Titrino

81

3.2.2.68. Info.ActualInfo.Assembly.Counter.V

read only

Info.ActualInfo.Assembly.Counter.Clear

$G

$Q sends the volume. With the function &Info.Assembly.Counter.Clear $G,
the volume counter is set to zero.


3.2.2.69. Info.ActualInfo.Assembly.Meas

read only

$Q sends the current measured value from the assembly.


3.2.2.70. Info.ActualInfo.Titrator.CyclNo

read only

Info.ActualInfo.Titrator.V

read only

Info.ActualInfo.Titrator.Meas

read only

Info.ActualInfo.Titrator.dVdt

read only

Info.ActualInfo.Titrator.dMeasdt

read only

Info.ActualInfo.Titrator.dMeasdV

read only

Info.ActualInfo.Titrator.T

read only

$Q sends the current values in the following formats:

KFT

CyclNo

127

V(mL)

1.2345

Meas:
Ipol (mV)
Upol (µA)


–345.6
–12.5

dVdt (µL/s)

2.5142

dMeasdt
Ipol (mV/s)
Upol (µA/s)


0.7957
0.7957

dMeasdV (mV/µL)

10.6326


NV: Not Valid.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.75). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.


3.2.2.71. Info.ActualInfo.MeasPt.Index

read only

Info.ActualInfo.MeasPt.X

read only

Info.ActualInfo.MeasPt.Y

read only

Info.ActualInfo.MeasPt.Z1

read only

Info.ActualInfo.MeasPt.Z2

read only

Info.ActualInfo.EP.Index

read only

Info.ActualInfo.EP.X

read only

Info.ActualInfo.EP.Y

read only

$Q sends the last entry into the measuring point list (.MeasPt) or the last en-
try into the list of EP's.