3.2 Remote control commands

784 KFP Titrino

83

3.2.2.76. Assembly.Bur.Rates.Forward.Selected

digital, analog

Assembly.Bur.Rates.Forward.Digital

0...150, max.

Assembly.Bur.Rates.Reverse.Selected

digital, analog

Assembly.Bur.Rates.Reverse.Digital

0...150, max.

Expel and aspirating rate.
Digital or analog control. With digital control, the inputted value applies (in
mL/min). "max." means maximum possible rate with the Exchange Unit in
current use.
Analog means rate control with the analog potentiometer on Titrino.


3.2.2.77. Assembly.Bur.Fill

$G, $H, $C

$G starts the 'FILL' mode of the burette function.


3.2.2.78. A
ssembly.Bur.ModeDis

$G, $S, $H, $C

Assembly.Bur.ModeDis.Selected

volume, time

Assembly.Bur.ModeDis.V

0.0001...0.1...9999

Assembly.Bur.ModeDis.Time

0.25...1...86400

Assembly.Bur.ModeDis.VStop

0.0001...9999, OFF

Assembly.Bur.ModeDis.AutoFill

ON, OFF

Dispensing mode with parameters. The dispensing mode can only be started
and stopped via the RS Control. During a running dosification, no method can
be started at the Titrino.
.Selected: Dispensing of volume increments or during a preset time.
.Volume, .Time: Size of the volume increments or entry of time.
.VStop:

Limit volume for the dispensing.

.AutoFill: ON means automatic filling after every dispensing.


3.2.2.79. Assembly.Meas.Status

ON, OFF

Assembly.Meas.MeasInput

Ipol, Upol

Assembly.Meas.Ipol

±127...1...+127

Assembly.Meas.Upol

±1270...400...+1270


Measurement in assembly. The measuring function can only be started via
RS Control. When the measuring function is switched on, no method can be
started at the Titrino.
.Input:

Selection of the "measuring input" for polarized electrodes.

.Ipol:

Polarization current in µA.

.Upol:

Polarization potential in mV, entry in steps of 10 mV.