3.2 Remote control commands
784 KFP Titrino
56
&Config
Object
Description
Input range
Reference
& Root
:
├
Config
Instrument configuration
:
├
.Monitoring
Monitoring functions
│ ├
.Validation
Validation monitoring
3.2.2.33.
│ │ ├
.Status
Status of validation monitoring
ON, OFF
ditto
│ │ ├
.Interval
Time interval for validation
1...365...9999
ditto
│ │ ├
.Counter
Time counter
0...9999
ditto
│ │ ├
.ClearCount
Clears the counter above
$G
ditto
│ ├
.Service
Monitoring of Metrohm service
$G
3.2.2.34.
│ │ ├
.Status
Status of service monitoring
ON, OFF
ditto
│ │ ├
.Date
Date of next service
YYYY-MM-DD
ditto
│ ├
.DiagRep
Printing of system test report
ON, OFF
3.2.2.35.
│
├
.PeriphUnit
Selection of peripheral units
│ ├
.CharSet1
External printer
Epson,Seiko,Citizen
│ │
IBM,HP
3.2.2.36.
│ ├
.CharSet2
as for CharSet1
│ ├
.RepToComport Output of manual reports
1, 2, 1&2
3.2.2.37.
│ ├
.Balance
Selection of balance
Sartorius,Mettler,Mettler AT
│ │
AND,Precisa
3.2.2.38.
│ ├
.Stirrer
Stirrer control
ON, OFF
3.2.2.39.
│ ├
.RemoteBox
Connected remote box
3.2.2.40.
│ │ ├
.Status
Status
ON, OFF
ditto
│ │ ├
.Keyboard
Type of keyboard
US, deutsch, francais,
│ │ │
español, schweiz.
ditto
│ │ ├
.Barcode
Input of barcode reader
input, method, id1, id2,
│ │ │
id3, smpl size
ditto
│
├
.Aux
Miscellaneous
│ ├
.Language
Dialog language
english, deutsch,
│ │
français, español, italiano,
│ │
portugese, svenska
3.2.2.41.
│ ├
.Set
Setting of date and time
$G
3.2.2.42.
│ │ ├
.Date
Date
YYYY-MM-DD
│ │ ├
.Time
Time
hh:mm
│ ├
.RunNo
Run number
0...9999
3.2.2.43.
│ ├
.AutoStart
Automatic start
1...9999, OFF
3.2.2.44.
│ ├
.StartDelay
Start delay time
0...999999
3.2.2.45.
│ ├
.ResDisplay
Result display
standard,bold
3.2.2.46.
│ ├
.DevName
Device label
8 ASCII char.
3.2.2.47.
│ ├
.Prog
Program version
read only
3.2.2.48.