Info – Metrohm 784 KFP Titrino User Manual
Page 65
3.2 Remote control commands
784 KFP Titrino
60
&Info
Object
Description
Input range
Reference
& Root
:
├
Info
Current data
:
├
.Report
Transmission of formatted reports $G
3.2.2.60.
│ ├
.Select
Report type
configuration, parameters,
│
smpl data, statistics, silo,
│
C-fmla, def, user method,
│
full, short, mplist, curve,
│
scalc full, scalc srt, calc,
│
all, ff
ditto
│
├
.Checksums
Checksums
$G
3.2.2.61.
│ ├
.MPList
Checksum of meas.point list
read only
ditto
│ ├
.ActualMethod
Checksum of current method
read only
ditto
│
├
.DetermData
Determination data
$G
3.2.2.62.
│ ├
.Write
Read/write for several nods
ON, OFF
│ ├
.ExV
Volume of Exchange/Dosing unit
read only/read+write
ditto
│ ├
.MPList
Measuring point list
│
│ ├
.1
Measuring point 1
│
│
│ ├
.Attribute Attribute
read only/read+write
ditto
│
│
│ ├
.X
X coordinate
read only/read+write
ditto
│
│
│ ├
.Y
Y coordinate
read only/read+write
ditto
│
│
│ ├
.Z1
Z1 coordinate
read only/read+write
ditto
│
│
│ ├
.Z2
Z2 coordinate
read only/read+write
ditto
│
│
│
for each measuring point
│
├
.TitrResults
Titration results
│ ├
.RS
Calculated results
│ │ ├
.1
1
st
result
│ │ │ ├
.Value
Value
read only
3.2.2.63.
│ │ ├
up to 9 results
│ ├
.EP
Endpoint
│ │ ├
.1
1
st
result
│
│ │ ├
.V
Value
read only
ditto
│
│ │ ├
.Meas
Measured value
read only
│
│ │ ├
.Mark
without meaning
│ │ ├
up to 9 EP's
│ ├
.Var
Variables C4X
│ │ ├
.C40
Start measured value
read only/read+write
ditto
│ │ ├
.C41
Titration end volume
read only/read+write
│ │ ├
.C42
Titration time
read only/read+write
│ │ ├
.C43
Volume drift
read only/read+write
│ │ ├
.C44
Titration temperature
read only/read+write
│ │ ├
.C45
Start volume
read only/read+write
│ │ ├
.DTime
Time for drift corr.or dosing time
read only/read+write