Metrohm 784 KFP Titrino User Manual
Page 78
3.2 Remote control commands
784 KFP Titrino
73
3.2.2.32. UserMeth.List.1.Name
read only
UserMeth.List.1.Mode
read only
UserMeth.List.1.Quantity
read only
UserMeth.List.1.DosUnit
read only
UserMeth.List.1.Bytes
read only
UserMeth.List.1.Checksum
read only
for each method
List of the methods in the user method memory with the following characte-
ristics:
.Name:
Name of the method
.Mode:
Mode
.Quantity: Measured quantity
.DosUnit: Buret of the method
.Bytes:
Number of bytes of the user memory used by the method
.Checksum: Checksum of the method, see 3.2.2.61.
3.2.2.33. Config.Monitoring.Validation.Status
ON, OFF
Config.Monitoring.Validation.Interval
1...365...9999
Config.Monitoring.Validation.Counter
0...9999
Config.Monitoring.Validation.ClearCount
$G
Monitoring of validation.
.Interval: Time interval in days for validation.
.Counter: Time counter in days since last validation.
.ClearCount: Clears the above counter.
3.2.2.34. Config.Monitoring.Service.Status
ON, OFF
Config.Monitoring.Service.Date
YYYY-MM-DD
Monitoring of service interval.
3.2.2.35. Config.Monitoring.DiagRep
ON, OFF
Printing of system test report after each switching on of the Titrino.
3.2.2.36. Config.PeriphUnit.CharSet1
Epson, Seiko, Citizen, HP, IBM
Config.PeriphUnit.CharSet2
Selection of the character set and the graphics control characters for COM1
resp. COM2 of the Titrino.
IBM means the IBM character set following character set table 437 and IBM
graphics control characters. Select 'IBM' for work with the computer.
3.2.2.37. Config.PeriphUnit.RepToComport
1, 2, 1&2
Selection of COM of the Titrino where manually triggered reports should be
outputted.
3.2.2.38. Config.PeriphUnit.Balance
Sartorius,Mettler,Mettler AT,
AND,Precisa
Selection of the balance type.