3.2 Remote control commands

784 KFP Titrino

54

*Parameter

Tree part "Parameters for KFT"

.

CtrlPara

Control parameters

│ ├

.EP

Endpoint

depends on meas.quant. 3.2.2.11.

│ ├

.UnitEp

Unit of endpoint

read only

ditto

│ ├

.Dyn

Dynamics

depends on meas.quant. 3.2.2.12.

│ ├

.UnitDyn

Unit of dynamics

read only

ditto

│ ├

.MaxRate

Maximum dosing rate

0.01...150, max.

ditto

│ ├

.MinIncr

Minimum increment

0.1...9.9, min.

ditto

│ ├

.Stop

Titration stop

│ │ ├

.Type

Type of stop criterion

drift, time

3.2.2.13.

│ │ ├

.Drift

Stop drift

1...20...999

ditto

│ │ ├

.Time

Switch-off delay time

0...10...999, inf

ditto

│ │ ├

.StopT

Stop time

0...999999, OFF

ditto

│ │

.

TitrPara

Titration parameters

│ ├

.Direction

Titration direction

+,

, auto

3.2.2.14.

│ ├

.XPause

Waiting time before start volume

0...999999

3.2.2.15.

│ ├

.StartV

Start volume

│ │ ├

.Type

Type of start volume

abs., rel., OFF

3.2.2.16.

│ │ ├

.V

Volume for absolute start volume

0...999.99

ditto

│ │ ├

.Factor

Factor for relative start volume

0...±999999

ditto

│ │ ├

.Rate

Dispensing rate for start volume

0.01...150.0, max.

ditto

│ ├

.Pause

Waiting time after start volume

0...999999

3.2.2.17.

│ ├

.ExtrT

Extraction time

0...999999

3.2.2.18.

│ ├

.MeasInput

without meaning

│ ├

.Ipol

Polarization current

0...50...±127

3.2.2.19.

│ ├

.Upol

Polarization voltage

0...400...±1270

ditto

│ ├

.PolElectrTest

Test for polarized electrodes

ON, OFF

ditto

│ ├

.Temp

Titration temperature

–170.0...25.0...500.0

3.2.2.20.

│ ├

.TDelta

Time interv. for meas.acquisition

1...2...999999

3.2.2.21.

│ │

.

StopCond

Stop conditions

│ ├

.VStop

Stop volume

│ │ ├

.Type

Type of stop volume

abs., rel., OFF

3.2.2.22.

│ │ ├

.V

Volume for absolute stop volume

0...99.99...9999.99

ditto

│ │ ├

.Factor

Factor for relative stop volume

0...±999999

ditto

│ ├

.FillRate

Filling rate

0.01...150.0, max.

3.2.2.23.

│ │

.

Statistics

Statistics

│ ├

.Status

Status of statistics calculation

ON, OFF

3.2.2.24.

│ ├

.MeanN

No. of individual determinations

2...20

ditto

│ ├

.ResTab

Result table

│ │ ├

.Select

original,delete n,delete all

ditto

│ │ ├

.DelN

Deletion of individual results

1...20

ditto

│ │

.

Presel

Preselections

│ ├

.Cond

Conditioning

ON, OFF

3.2.2.25.

│ ├

.DriftDisp

Display of drift during cond.

ON, OFF

ditto

│ ├

.DCor

Drift correction

│ │ ├

.Type

Type of drift acquisition

auto, man., OFF

ditto

│ │ ├

.Value

Drift value for manual drift corr.

0.0...99.9

ditto

│ ├

.IReq

Request of Id's after start

id1, id1&2, all, OFF

3.2.2.26.

│ ├

.SReq

Request of smpl size after start

value, unit, all, OFF

ditto

│ ├

.LimSmplSize

Limits for sample size

3.2.2.27.

│ │ ├

.Status

Status of limit control

ON, OFF

ditto

│ │ ├

.LoLim

Lower limit

0.0...999 999

ditto

│ │ ├

.UpLim

Upper limit

0.0...999 999

ditto

│ ├

.Oven

KF Oven connected

COM1, COM2, no

3.2.2.28.

│ ├

.ActPulse

Output of a pulse

first, all, cond., OFF

3.2.2.29.