beautypg.com

Metrohm 784 KFP Titrino User Manual

Page 73

background image

3.2 Remote control commands

784 KFP Titrino

68

The limit control for results can also be activated. If a result is out of limit, a
message appears in the result report, E196 is sent, and output line L13 can
be set.


3.2.2.6.

Mode.Def.SiloCalc.Assign.C24

RSX, EPX, CXX

Mode.Def.SiloCalc.Assign.C25

RSX, EPX, CXX

Mode.Def.SiloCalc.MatchId

id1, id1&2, all, OFF

.Assign.C2X: Assignment to store results in the silo as C2X.
.MatchId:

Indication which sample identification(s) have to match so that
the results can be combined.



3.2.2.7.

Mode.Def.ComVar.C30

RSX, MNX, EPX, CXX

Mode.Def.ComVar.C31

etc., up to .C39

Assignment of common variables.
The values of the common variables are to be found in &Config.ComVar. They
can be viewed and entered there, see 3.2.2.50.


3.2.2.8.

Mode.Def.Report.Assign1

Mode.Def.Report.Assign2

param, full, short, mplist, curve, scalc full, scalc srt, calc, ff

Definition of the report sequence, which is outputted automatically at the end
of the determination. Entries of more than one block have to be separated
with ";".
.Assign1: Output to COM1 of the Titrino. Identical for COM2.


3.2.2.9.

Mode.Def.Mean.1.Assign

RS1, RSX, EPX, CXX

Mode.Def.Mean.2.Assign

etc., up to .9

Assignment of the statistics calculations. Valid assignments are a require-
ment for statistics calculations. In addition, the statistics calculation must be
switched on, see 3.2.2.24. Rules for statistics calculations see page

29.



3.2.2.10. Mode.CFmla

Mode.CFmla.1.Value

0...±999999

Mode.CFmla.2.Value

etc., up to .19

Calculation constants specific to a method. Stored in the method memory of
the Titrino. Operands specific to the sample (3.2.2.52 and 3.2.2.63) and val-
ues of common variables (3.2.2.50) on the other hand are not stored with the
methods.