3.2 Remote control commands
784 KFP Titrino
70
3.2.2.16. Mode.Parameter.TitrPara.StartV.Type
abs., rel., OFF
Mode.Parameter.TitrPara.StartV.V
0...999.99
Mode.Parameter.TitrPara.StartV.Factor
0...±999999
Mode.Parameter.TitrPara.StartV.Rate
0.01...150, max.
Start volume.
If an absolute
A
start volume (abs.) has been selected, the volume in mL is va-
lid.
relative
Start volume in mL = smpl size ∗ factor
start volume (rel.) is dispensed as a function of the sample size:
The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.
3.2.2.17. Mode.Parameter.TitrPara.Pause
0...999999
Pause time in s. Is waited off after the dispensing of the start volume.
3.2.2.18. Mode.Parameter.TitrPara.ExtrT
0...999999
Extraction time in s.
3.2.2.19. Mode.Parameter.TitrPara.Ipol
–127...50...+127
Mode.Parameter.TitrPara.Upol
–1270...400...+1270
Mode.Parameter.TitrPara.PolElectrTest
ON, OFF
With Ipol, the inquiries for the polarization current in µA (Ipol) and
.PolElectrTest are valid.
With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry
in steps of 10 mV.
Besides .PolElectrTest is valid.
If the test for polarized electrodes is switched on, it is performed on chan-
geover from the inactive state to an active state (titration or conditioning).
3.2.2.20. Mode.Parameter.TitrPara.Temp
–170.0...25.0...500.0
Titration temperature in °C.
3.2.2.21. Mode.Parameter.TitrPara.TDelta
1...2...999999
Time interval in s for the entry of a measurement point in the list of measured
points.