3.2 Remote control commands

784 KFP Titrino

70

3.2.2.16. Mode.Parameter.TitrPara.StartV.Type

abs., rel., OFF

Mode.Parameter.TitrPara.StartV.V

0...999.99

Mode.Parameter.TitrPara.StartV.Factor

0...±999999

Mode.Parameter.TitrPara.StartV.Rate

0.01...150, max.

Start volume.
If an absolute

A

start volume (abs.) has been selected, the volume in mL is va-

lid.

relative

Start volume in mL = smpl size ∗ factor

start volume (rel.) is dispensed as a function of the sample size:

The factor is valid.
The dispensing rate in mL/min applies to both cases. Max. means maximum
possible dispensing rate with the Exchange Unit in current use.


3.2.2.17. Mode.Parameter.TitrPara.Pause

0...999999

Pause time in s. Is waited off after the dispensing of the start volume.


3.2.2.18. Mode.Parameter.TitrPara.ExtrT

0...999999

Extraction time in s.


3.2.2.19. Mode.Parameter.TitrPara.Ipol

–127...50...+127

Mode.Parameter.TitrPara.Upol

–1270...400...+1270

Mode.Parameter.TitrPara.PolElectrTest

ON, OFF

With Ipol, the inquiries for the polarization current in µA (Ipol) and
.PolElectrTest are valid.
With Upol, the inquiry for the polarization voltage in mV (Upol) is valid. Entry
in steps of 10 mV.
Besides .PolElectrTest is valid.
If the test for polarized electrodes is switched on, it is performed on chan-
geover from the inactive state to an active state (titration or conditioning).


3.2.2.20. Mode.Parameter.TitrPara.Temp

–170.0...25.0...500.0

Titration temperature in °C.


3.2.2.21. Mode.Parameter.TitrPara.TDelta

1...2...999999

Time interval in s for the entry of a measurement point in the list of measured
points.