3.1 General rules
784 KFP Titrino
50
RS receive errors:
E36
Parity.
Exit:
both devices are the same.
E37
Framing error.
Exit:
both devices are the same.
E38
Overrun error. At least 1 character could not be read.
Exit:
E39
The internal working-off buffer of the Titrino is full
(>82 characters).
Exit:
RS send errors:
E42
CTS=OFF No proper handshake for more than 1 s.
Exit:
E43
The transmission of the Titrino has been interrupted with XOFF for
at least 6 s.
Exit: Send XON or
E45
The receive buffer of the Titrino contains an incomplete command
(L
F
missing). Sending from the Titrino is therefore blocked.
Exit: Send L
F
or
E121
Measuring point list overflow (more than 500 measuring points).
Exit: The error message disappears on next startup.
E123
Missing EP for calculation.
Exit: The error message disappears on next startup or on
recalculation.
E128
No new mean.
Exit: The error message disappears on next startup or on
recalculation.
E129
No new common variable, old value remains.
Exit: The error message disappears on next startup or on
recalculation.
E130
Wrong sample. For KFT with preset titration direction the first
measured value lies behind the endpoint.
Exit: The error message disappears on next startup.
E132
Silo empty and it has been started with open silo or empty silo has
been opened.
Exit: Send a silo entry.
E133
Silo full.
Exit: Send new command.
E137
XXX Bytes are missing so that the method or the silo line could not be stored.
Exit: Send new command.
E155
No new silo result (C24 or C25).
Exit: The error message disappears on next start or on recalculation.