3.2 Remote control commands

784 KFP Titrino

86

3.2.2.87. Setup.Mode.StartWait

ON, OFF

Setup.Mode.FinWait

ON, OFF

Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method (holding point after Au-

toInfo !".T.GC").

.FinWait: Holding point at the end a method (holding point after AutoInfo

!".T.F").



3.2.2.88. Setup.SendMeas.SendStatus

ON, OFF

Setup.SendMeas.Interval

0.08...4...16200, MPList

.SendStatus: ON means the automatic transmission of measured values (see

3.2.2.90 and 3.2.2.91) in the inputted interval is active.

.Interval: Time interval (in s) for the automatic transmission of associated

measured values defined under points 3.2.2.90 and 3.2.2.91. The
inputted value is rounded off to a multiple of 0.08. The smallest
possible time interval depends on the number of measured values
which have to be sent, on the baud rate, on the load on the inter-
face and on the type of device connection. With "MPList" the
measured values are sent at the time of their entry into the meas-
ured point list.

The automatic transmission is switched on/off with 'SendStatus'.


3.2.2.89. Setup.SendMeas.Select

Assembly, Titrator

Selection of the unit of which the measured values should be sent (3.2.2.90
and 3.2.2.91).


3.2.2.90. Setup.SendMeas.Assembly.CyclNo

ON, OFF

Setup.SendMeas.Assembly.V

ON, OFF

Setup.SendMeas.Assembly.Meas

ON, OFF

Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.88):
.CyclNo: Cycle number of the potential measurement. Together with the

cycle time (3.2.2.75), a time frame can be set up.

The cycle number is set to 0 on switching on the instrument and
it is always incremented as long as the instrument remains
switched on.

.V:

Volume

.Meas:

Measured value associated to the cycle number.

The unit "assembly" must be preset (see 3.2.2.89).


3.2.2.91. Setup.SendMeas.Titrator.CyclNo

ON, OFF

Setup.SendMeas.Titrator.V

ON, OFF

Setup.SendMeas.Titrator.Meas

ON, OFF

Setup.SendMeas.Titrator.dVdt

ON, OFF

Setup.SendMeas.Titrator.dMeasdt

ON, OFF

Setup.SendMeas.Titrator.dMeasdV

ON, OFF