Metrohm 784 KFP Titrino User Manual
Page 91
3.2 Remote control commands
784 KFP Titrino
86
3.2.2.87. Setup.Mode.StartWait
ON, OFF
Setup.Mode.FinWait
ON, OFF
Holding points in the method sequence. If they are "ON", the sequence stops
until "OFF" is sent. Switching the instrument on sets both nodes to OFF:
.StartWait: Holding point right after starting a method (holding point after Au-
toInfo !".T.GC").
.FinWait: Holding point at the end a method (holding point after AutoInfo
!".T.F").
3.2.2.88. Setup.SendMeas.SendStatus
ON, OFF
Setup.SendMeas.Interval
0.08...4...16200, MPList
.SendStatus: ON means the automatic transmission of measured values (see
3.2.2.90 and 3.2.2.91) in the inputted interval is active.
.Interval: Time interval (in s) for the automatic transmission of associated
measured values defined under points 3.2.2.90 and 3.2.2.91. The
inputted value is rounded off to a multiple of 0.08. The smallest
possible time interval depends on the number of measured values
which have to be sent, on the baud rate, on the load on the inter-
face and on the type of device connection. With "MPList" the
measured values are sent at the time of their entry into the meas-
ured point list.
The automatic transmission is switched on/off with 'SendStatus'.
3.2.2.89. Setup.SendMeas.Select
Assembly, Titrator
Selection of the unit of which the measured values should be sent (3.2.2.90
and 3.2.2.91).
3.2.2.90. Setup.SendMeas.Assembly.CyclNo
ON, OFF
Setup.SendMeas.Assembly.V
ON, OFF
Setup.SendMeas.Assembly.Meas
ON, OFF
Selection of the values from Assembly for the output in the set time interval
(see 3.2.2.88):
.CyclNo: Cycle number of the potential measurement. Together with the
cycle time (3.2.2.75), a time frame can be set up.
The cycle number is set to 0 on switching on the instrument and
it is always incremented as long as the instrument remains
switched on.
.V:
Volume
.Meas:
Measured value associated to the cycle number.
The unit "assembly" must be preset (see 3.2.2.89).
3.2.2.91. Setup.SendMeas.Titrator.CyclNo
ON, OFF
Setup.SendMeas.Titrator.V
ON, OFF
Setup.SendMeas.Titrator.Meas
ON, OFF
Setup.SendMeas.Titrator.dVdt
ON, OFF
Setup.SendMeas.Titrator.dMeasdt
ON, OFF
Setup.SendMeas.Titrator.dMeasdV
ON, OFF