Metrohm 813 Compact Autosampler User Manual
Page 39
4
Appendix
813 Compact Autosampler
33
Method 3
Double determination of each sample, no air gap, needle remains
in sample after sample transfer.
813 Compact Autosampler 813.0010
parameters
method 813_3
number of samples: *
>start sequence
1 SAMPLE: = 1
>sample sequence
1 SCN:Rm: *****1**
2 MOVE 1 : sample
3 LIFT: 1 : work mm
4 SCN:Rm: *****1**
5 PUMP 1.1 : on
6 WAIT 5 s
7 SCN:Rm : *****1**
8 WAIT 12 s
9 PUMP 1.1 : off
10 SCN:Rm: *****1**
11 WAIT 5 s
12 SCN:Rm: *****1**
13 PUMP 1.1 : on
14 WAIT 5 s
15 SCN:Rm : *****1**
16 WAIT 12 s
17 PUMP 1.1 : off
>final sequence
>changer settings
rack number 0
lift rate 12 mm/s
shift rate 20
shift direction: auto.
>manual stop
------------
¬
Program version
¬
Method name
¬
Number of samples unlimited
¬
First rack position
¬
Await 1
st
Remote signal
¬
Move next sample to tower
¬
Lift to work position / immerse needle
¬
Await 2
nd
Remote signal
¬
Switch on pump
¬
Waiting time
¬
Await 3
rd
Remote signal
¬
Waiting time for 1
st
sample transfer
¬
Switch off pump
¬
Await 1
st
Remote signal again
¬
Waiting time
¬
Await 2
nd
Remote signal
¬
Switch on pump
¬
Waiting time
¬
Await 3
rd
Remote signal
¬
Waiting time for 2
nd
sample transfer
¬
Switch off pump
------ Settings for changer functions --------------