beautypg.com

Metrohm 813 Compact Autosampler User Manual

Page 38

background image

4.2 Sequences

32

813 Compact Autosampler

Method 2

Single determination, no air gap, needle remains in sample after

sample transfer.

813 Compact Autosampler 813.0010
parameters
method 813_2
number of samples: *
>start sequence
1 SAMPLE: = 1
>sample sequence
1 SCN:Rm: *****1**
2 MOVE 1 : sample
3 LIFT: 1 : work mm
4 SCN:Rm: *****1**
5 PUMP 1.1 : on
6 WAIT 5 s
7 SCN:Rm : *****1**
8 WAIT 12 s
9 PUMP 1.1 : off
>final sequence
>changer settings
rack number 0
lift rate 12 mm/s
shift rate 20
shift direction: auto.
>manual stop
------------

¬

Program version


¬

Method name

¬

Number of samples unlimited


¬

First rack position


¬

Await 1

st

Remote signal

¬

Move next sample to tower

¬

Lift to work position / immerse needle

¬

Await 2

nd

Remote signal

¬

Switch on pump

¬

Waiting time

¬

Await 3

rd

Remote signal

¬

Waiting time for sample transfer

¬

Switch off pump


------ Settings for changer functions --------------