2.3 Attaching the accessories
16
813 Compact Autosampler
2.3.9 Tubing connection to a Metrohm 747 VA Stand
3 Connection Pumpschlauch – VA Stand 747
• Use a 6.2744.010 PEEK compression fitting and a 6.2744.030
coupling to tighten the 6.1831.060 PEEK capillary
17
to the
pump tubing
8
.
To keep the dead volume as small as possible shorten the
pump tubing to a suitable length.
4 Measuring head of the 747 VA Standes
• Insert the other end of the PEEK capillary
17
into a threaded
opening of the Measuring head of the VA Stand and fix it
with a 4.420.2580 fitting screw.
Metrohm
747
Fig. 10 Connection of a 747 VA Stand
Fig. 11 Measuring head of the747 VA Stand
2.3.10
Tubing connection to a Metrohm Computrace 757
Install tubing connections 813 Compact Autosampler – 757 VA
Computrace as shown above for the 747 VA Stand. See the 757
VA Computrace user manual for more details.
sample
6.1822.410 PEEK capillary
rinsing
waste
elektrolyte
gassing /
vent
4-way-micro tip
- 915 KF Ti-Touch (382 pages)
- 800 Dosino (53 pages)
- 767 Calibrated Reference (23 pages)
- 940 Professional IC Vario ONE/SeS/Prep 2 (54 pages)
- 754 Dialysis Unit (49 pages)
- 815 Robotic Soliprep for LC (76 pages)
- Vision Manual (207 pages)
- tiamo 2.1 Manual (1532 pages)
- 825 Lab Link (37 pages)
- 808 Titrando (70 pages)
- 902 Titrando (52 pages)
- 756 KF Coulometer (162 pages)
- 756 KF Coulometer (163 pages)
- 940 Professional IC Vario ONE/LPG (98 pages)
- 850 Professional IC Anion MCS Prep 3 (154 pages)
- 850 Professional IC Anion MCS Prep 3 (152 pages)
- 904 Titrando (58 pages)
- 850 Professional IC Anion MSM-HC MCS Prep 2 (150 pages)
- 930 Compact IC Flex Oven/ChS/Deg (47 pages)
- 872 Extension Module Liquid handling (64 pages)
- 814 USB Sample Processor (90 pages)
- 814 USB Sample Processor (91 pages)
- 940 Professional IC Vario (43 pages)
- Vision – Tutorial (40 pages)
- 799 GPT Titrino (242 pages)
- 889 IC Sample Center (68 pages)
- 761 Compact IC (228 pages)
- 851 Titrando (100 pages)
- 748 DH Sample Changer (32 pages)
- 940 Professional IC Vario ONE/SeS/HPG (51 pages)
- 896 Professional Detector – Amperometry (62 pages)
- 877 Titrino plus (139 pages)
- 881 Compact IC pro – Anion (129 pages)
- 940 Professional IC Vario ONE/ChS/HPG (112 pages)
- 930 Compact IC Flex Deg (41 pages)
- 840 PC Control 5.0 / Touch Control (351 pages)
- 940 Professional IC Vario ONE/Prep 1 (45 pages)
- 776 Dosimat (42 pages)
- 717 Sample Changer (36 pages)
- 815 Robotic USB Sample Processor XL (113 pages)
- 815 Robotic USB Sample Processor XL (114 pages)
- 940 Professional IC Vario ONE/SeS/PP (126 pages)
- 838 Advanced Sample Processor Installation Instructions (109 pages)
- 700 Dosino (55 pages)
- 719 S Titrino (152 pages)