4.2 Sequences
The available Autosampler methods are described briefly in sec-
tion 3.4 The following listings explain the sequences in detail.
Method 1
Single determination with air gap after each run, needle lifted after
sample transfer.
813 Compact Autosampler 813.0010
parameters
method 813_1
number of samples: *
>start sequence
1 SAMPLE: = 1
>sample sequence
1 SCN:Rm: *****1**
2 MOVE 1 : sample
3 LIFT: 1 : work mm
4 SCN:Rm: *****1**
5 PUMP 1.1 : on
6 WAIT 5 s
7 SCN:Rm : *****1**
8 WAIT 12 s
9 PUMP 1.1 : off
10 LIFT: 1 : 0 mm
11 PUMP 1.1 : 6 s
>final sequence
>changer settings
rack number 0
lift rate 12 mm/s
shift rate 20
shift direction: auto.
>manual stop
------------
¬
Program version
¬
Method name
¬
Number of samples unlimited
¬
First rack position
¬
Await 1
st
Remote signal
¬
Move next sample to tower
¬
Lift to work position / immerse needle
¬
Await 2
nd
Remote signal
¬
Switch on pump
¬
Waiting time
¬
Await 3
rd
Remote signal
¬
Waiting time for sample transfer
¬
Switch off pump
¬
Lift to rest position / raise needle
¬
Switch on pump for 6 seconds
------ Settings for changer functions --------------