2.3.5 Placing the sample rack

1

2

Place sample rack

Place sample rack

24

on the

turntable of the 813 Compact

Autosampler acc. to Fig. 8.

Initialize rack

Press to move the

rack to the home position, in

which the magnetic code can

be read (coding magnet 30).

30

Fig. 8 Sample rack placing

2.3.6 Adjusting the sample rack

If a new sample rack is placed on the 813 Compact Autosampler

for the first time, it must be adjusted. Make sure you have

mounted a steel needle or a PEEK needle before. Proceed as fol-

lows:

1 Place sample rack

Place empty sample rack

24

on the turntable of the 813 Com-

pact Autosampler and switch off the instrument.

2 Switch on the instrument

Keep the key pressed and simultaneously press the

mains switch on the right hand side of the

813 Compact Auto-

sampler. The instrument is initialized and the needle is placed

above rack position 1. Press the key. The green LED