beautypg.com

Metrohm 813 Compact Autosampler User Manual

Page 19

background image

2

Installation

813 Compact Autosampler

13

3 Check needle position

• Press <

HOLD

> until the needle

2

is about 1 cm above the

upper level of the sample rack

24

.

• Check needle position: If the needle 2 cannot be lowered

unhindered through the upper hole of the rack position 1,

continue directly with point 4.

• Press once and continue lowering the needle by

pressing <

HOLD

> again until the needle

2

stops.

• Check needle position: If the needle

2

points to the middle of

lower hole of the rack position 1, the sample rack must not be

adjusted (continue directly with point 5).

4 Adjusting the sample rack

• Loosen the four adjusting screws

33

on the lower level of the

sample rack using the 6.2621.100 Allen key .

• Carefully turn the two upper levels of the sample rack

27

by

hand until the lowered needle

2

is exactly in the middle of the

lower hole at the rack position 1.

• Tighten the adjusting screws

33

.

5 Move to rest position

Press twice to move the sample rack to the initial

position.