3.3 RAM initialization
24
813 Compact Autosampler
4. Accept the selected method with
5. To return to the basic state, press the
If you just want to define the work position:
1. Switch off Autosampler .
2. Press
The green LED blinks.
3. Press
4. Press
slowly. Press
sition exactly. Each time the
rection of the lift movement is reversed. The PEEK needle
should be immersed deep enough to assure a trouble-free
aspiration of the sample or. Depending on your application
you may want to aspirate tho whole sample. In the latter case
drive the lift down until the needle reaches the bottom of the
sample vial.
5. The lift position set can be accepted with the
6. Reestablish basic state of the 813 Compact Autosampler by
pressing the
3.3 RAM initialization
The internal software of the 813 Compact Autosampler may be
initialized to reset all instrument settings after serious malfunc-
tions, which should be very scarce of course. All settings (includ-
ing the not accesible ones) will be reset to factory defaults.
Initialization can be done by pressing all keys simultaneously
while switching on the instrument.
It may be appropriate to initialize the 813 Compact Autosampler
before using it the first time.
START
HOLD
STOP
RESET
and "Power on"
Method selection and the definition of the working position has to
be redone after RAM initialization.