6 modular metrohm ic system 732/733, 7 metrohm 746 va trace analyzer, 1 monitoring – Metrohm 813 Compact Autosampler User Manual
Page 34: 2 command sequence
![background image](/manuals/586261/34/background.png)
3.6 Modular Metrohm IC System 732/733
28
813 Compact Autosampler
3.6 Modular Metrohm IC System 732/733
The 813 Compact Autosampler can be controlled directly by a
Metrohm 732 IC Detector or by a PC control software, such as the
Metrohm IC Net 2.0 using the remote control lines of the 732 IC
Detector. See the 732 IC Detector user manual for details or read
the IC Net 2.0 documentation.
3.7 Metrohm 746 VA Trace Analyzer
Using a 813 Compact Autosamplers with a 746 VA Trace Ana-
lyzer require modified command sequences and monitoring set-
tings.
Between each sample determination a rinsing procedure for the
tubings has to be executed. For this purpose place a sample
tube containing rinsing solution on each second (even) rack posi-
tion. The samples to be determined have to be placed between
those rinsing tubes. This reduces the amount of a sample series
to 18 samples max.
3.7.1 Monitoring
The number of samples of a series is defined on the monitoring
page of the 746 VA Trace Analyzer, see below.
Auto.samples on 18 of 160
Auto.batch on 0 of 18
Start interval 0.0 h
3.7.2 Command sequence
A command sequence of the 746 VA Trace Analyzer has to con-
tain explicit trigger signals for the 813 Compact Autosampler that
are to be sent via the Remote output line 2. Use the 'SETCTRL'
command to send the required pulses, see line 4 and 5 of the ex-
ample method on the next page.