10.12 Temperature variations

Temperature variations can significantly influence the measuring results. To reduce this
influence, it is recommended to check and store the probe constant regularly.

11 Maintenance

11.1 Cleaning

The plastic parts, the display unit as well as the painted parts of the instrument should be cleaned using a
slightly wet (watered) fabric. Clean the air cushion pads using a clean fabric, slightly soaked with alcohol.


12 Technical specifications

Mestra 300

Mestra 600

Mestra 1000

Measuring range

mm

305

610

1016

Application range

mm

567

872

1278

Max. resolution

mm

0.0001

Max. permissible error

µm

1.4 + L(mm)/400

Repeatability

µm

0.5 (diameters:1)

Max. manual displacement speed

mm/s

1000

Max. motorized displacement speed

mm/s

150

Measuring force

N

0.5 .. 1.8

Power supply

Incorporated battery pack, rechargeable

Autonomy h

>

8

Overall squareness deviation using probe

µm

4 6 10

Data output

2 x RS232 C et 2 x USB (A et B)

Operational temperature

°C

+10 .. +40

Storage temperature

°C

-10 .. +40

Total height

mm

645

950

1357

Weight kg

16

19

23

The values of max. permissible error and repeatability are valid only when using the standard probe (measuring insert with tungsten carbide
ball,

∅ 4 mm, L = 90 mm) at temperature of 20 ± 0.5 °C and relative humidity of 50 ±5%.



13 After sales service

13.1 Complaints / repairs

In case of problems, please contact your local TRIMOS agent.
For any transport, use the original packing or an adequate one.

The warranty is only valid when it has been checked and signed by the TRIMOS agent.

13.2 Agents

You can find the official TRIMOS agents list on the Internet site : www.trimos.ch