10.12 Temperature variations
Temperature variations can significantly influence the measuring results. To reduce this
influence, it is recommended to check and store the probe constant regularly.
11 Maintenance
11.1 Cleaning
The plastic parts, the display unit as well as the painted parts of the instrument should be cleaned using a
slightly wet (watered) fabric. Clean the air cushion pads using a clean fabric, slightly soaked with alcohol.
12 Technical specifications
Mestra 300
Mestra 600
Mestra 1000
Measuring range
mm
305
610
1016
Application range
mm
567
872
1278
Max. resolution
mm
0.0001
Max. permissible error
µm
1.4 + L(mm)/400
Repeatability
µm
0.5 (diameters:1)
Max. manual displacement speed
mm/s
1000
Max. motorized displacement speed
mm/s
150
Measuring force
N
0.5 .. 1.8
Power supply
Incorporated battery pack, rechargeable
Autonomy h
>
8
Overall squareness deviation using probe
µm
4 6 10
Data output
2 x RS232 C et 2 x USB (A et B)
Operational temperature
°C
+10 .. +40
Storage temperature
°C
-10 .. +40
Total height
mm
645
950
1357
Weight kg
16
19
23
The values of max. permissible error and repeatability are valid only when using the standard probe (measuring insert with tungsten carbide
ball,
∅ 4 mm, L = 90 mm) at temperature of 20 ± 0.5 °C and relative humidity of 50 ±5%.
13 After sales service
13.1 Complaints / repairs
In case of problems, please contact your local TRIMOS agent.
For any transport, use the original packing or an adequate one.
The warranty is only valid when it has been checked and signed by the TRIMOS agent.
13.2 Agents
You can find the official TRIMOS agents list on the Internet site : www.trimos.ch