750 50 0001 03
18/28
3. To activate a tolerance input mode, press the Enter key.
The fields at the right side of the symbol will than be
activated.
To cancel the mode, press the Enter key again.
4. Recall the active fields using the cursor keys. Enter the
nominal size and the corresponding tolerance limits
(numerical key pad).
5. Once the values entered, press the Enter key longer
than 2 seconds to quit menu.
> 2 sec.
6.10.2 Application of the tolerance limits mode
1. To activate the tolerance mode, press the Tol key. The
text "Tol" appears above the current measuring value.
To cancel this mode, press the Tol key again.
Note:
If no tolerance has been entered, it is impossible to
activate this mode.
2. For each surface, diameter or centerline distance
measurement of which the tolerance limits mode has
been activated, luminous LED's will indicate the
tolerance position.
Red LED :
Out of tolerances,
rework of the part not possible
Green LED :
In tolerances
Orange LED : Out of tolerances,
possibility to rework the part
6.11 Complete clearing of the buffer
To clear the contents of the buffer completely, press the
Clear buffer key longer than 2 seconds.
> 2 sec.
6.12 Clearing of the last displayed value in the buffer
After each measurement, the value will be stored in the
buffer. To clear always the last value, press the Clear key.
Red
Green Orange