67 • Using Maintenance Mode
10,000 ppm Gas Tracer Operator’s Manual
•
To return to the BUMP screen in PPM CAL, press POWER
ENTER at any time. If the sensor fails calibration, the buzzer
will sound a double pulsing tone until you press and release
the POWER ENTER button.
•
Disconnect the humidifier from the probe.
NOTE:
Grasp the humidifier tube assembly by the clear tubing at the
end, not the humidifier tube itself.
11.Unscrew the demand flow regulator from the calibration cylinder.
12.Use the (SHIFT)▼ button to navigate to the ESCAPE menu item,
then press and release the POWER ENTER button to return to the
PPM CAL
menu in Calibration Mode.
13.Use the (SHIFT)▼ button to navigate to the NORMAL menu item,
then press and release the POWER ENTER button to return to
Measuring Mode.
Turning the Lunch Break Setting On or Off
The Lunch Break Setting does not have any effect on the operation of
the Gas Tracer.
Updating the Alarm Latching Setting
The Alarm Latching Setting does not have any effect on the operation
of the Gas Tracer.