10,000 ppm Gas Tracer Operator’s Manual
Start Up • 18
•
If calibration is not due or CL EXPRD is set to NO EFFECT,
then the following screen appears for a few seconds indicating
when the next calibration is due.
5. If BP RMNDR is turned on (factory setting is OFF), and
B--CHECK
is set to ALL (factory setting), the screen that appears
next depends on how BP EXPRD is set in the Maintenance Mode
Menu (page 72).
•
If the unit is due for a bump test and BP EXPRD is set to
CONFIRM (factory setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After
this, press the RESET SILENCE button until you hear a beep to
continue.
•
If the unit is due for a bump test and BP EXPRD is set to NOT
USE, then the following screen displays.
The Gas Tracer cannot be used until a bump test has been
performed by selecting PPM CAL in the Maintenance Mode
menu and then selecting BUMP. See “Performing a Bump
2011
9
30
69d
NEXT CAL
tESt
B--LIMIT
FAIL
B--LIMIT