Cal c--limit fail c--limit – RKI Instruments Gas Tracer 10,000 ppm User Manual
Page 24

17 • Start Up
10,000 ppm Gas Tracer Operator’s Manual
3. Press and briefly hold down the POWER ENTER button.
4. If CL RMNDR is turned on (factory setting is OFF), and
C--CHECK
is set to ALL (factory setting), the screen that appears
next depends on how CL EXPRD is set in the Maintenance Mode
Menu (page 72).
•
If the unit is due for calibration and CL EXPRD is set to
CONFIRM (factory setting), then the following screen displays.
The alarm LED’s and buzzer will pulse several times. After
this, press and release the RESET SILENCE button to continue.
•
If the unit is due for calibration and CL EXPRD is set to NOT
USE, then the following screen displays.
The Gas Tracer cannot be used until a calibration has been
performed either by selecting PPM CAL in the Maintenance
Mode menu. See “Calibrating and Bump Testing the PPM
Combustible Sensor” on page 54 for calibration instructions.
CAL
C--LIMIT
FAIL
C--LIMIT
- Eagle Series (89 pages)
- Eagle Series (66 pages)
- 01 Series Manual (33 pages)
- 01 Series Quik Start (2 pages)
- OX-94 (13 pages)
- FP-3 (16 pages)
- GasWatch 1 (18 pages)
- GasWatch 2 Manual (37 pages)
- GasWatch 2 Quik Start (2 pages)
- GP-01 (34 pages)
- GP-01 with self resetting alarms (33 pages)
- GP-204 (7 pages)
- NP-204 (13 pages)
- RI-215A (24 pages)
- SC-01 (68 pages)
- SC-01 (45 pages)
- Data Cal 2000 (88 pages)
- EAGLE 2 Manual (309 pages)
- Eagle 2 Quik Start (2 pages)
- GX-94 (3 pages)
- GX-94 (52 pages)
- SDM-E2 (224 pages)
- SDM-E2 (89 pages)
- Eagle 1 Manual (89 pages)
- Eagle 1 Quik Start (2 pages)
- Inert Gas Testing Eagle with Two Pumps (66 pages)
- Gas Tracer (234 pages)
- GX-2001 Manual (45 pages)
- GX-2001 Quik Start (2 pages)
- GX-2001 Canadian Version (43 pages)
- GX-2003 Manual (94 pages)
- GX-2003 Quik Start (2 pages)
- GX-2009 Manual (73 pages)
- GX-2009 Quik Start (2 pages)
- GX-2009 Supplement (2 pages)
- GX-2012 Manual (185 pages)
- GX-82 (22 pages)
- GX-82A (43 pages)
- GX-86 (22 pages)
- GX-86A (39 pages)
- GX-8000(PT0-098) (52 pages)
- GX-8000(H4-0050) (23 pages)
- RX-516 (36 pages)
- GX-2009 (1 page)