Bump, Bump 30, Performing a bump test – RKI Instruments Gas Tracer 10,000 ppm User Manual
Page 115

10,000 ppm Gas Tracer Operator’s Manual
Calibrating and Bump Testing the PPM Com-
Performing a Bump Test
This section describes how to bump test the PPM combustible gas
sensor.
NOTE:
The BUMP DSP menu item in Maintenance Mode is factory
set to OFF. The BUMP menu item will not appear unless
BUMP DSP
is set to ON. See “Turning the Bump Test
Function On or Off” on page 68 for instructions.
NOTE:
The Gas Tracer uses parameters defined in the BUMP--SET
menu item of Maintenance Mode while performing a bump
test. To view or update these parameters, see “Updating the
Bump Test Parameters” on page 69.
1. Use the AIR▲ or (SHIFT)▼ buttons to display the BUMP menu
item.
2. Fill the gas bag with ppm calibration gas as described in “Filling
the Gas Bag” on page 98.
3. Press and release the POWER ENTER button. You will be
prompted to choose between WET and DRY. Use the WET setting.
NOTE:
The DRY setting is for factory use only. Do not perform a
bump test using the DRY setting.
4. Use the AIR▲ or (SHIFT)▼ buttons to display the WET setting and
press and release the POWER ENTER button. The bump test gas
value will be displayed.
MAINTENANCE
BUMP
1000
ppm
MAINTENANCE
CH
4
BUMP 30