MTS Landmark System User Manual
Page 80

a) To open the Tuning: Ch1 Force window, select Station Manager > Display > Station Setup >
Channels > Ch1 (or other depending on your configuration) > Force and then click the Tuning
Fork icon.
b) Review the following parameters for this hard specimen test. The following is a good place to start
for a HCF test with a hard specimen.
•
P Gain — 1
•
I Gain — 1
•
D Gain — 0
•
F Gain — 0
•
F2 Gain — 0
•
Proportional FF Gain — 0
•
FL Filter — Set to maximum positive value.
3. Disable manual command.
This allows the function generator to control the system.
a) In the Station Manager main window, click the Manual Command button. The Manual Commands
window appears.
b) Click to clear (uncheck) the Enable Manual Command check box.
4. Set up a force waveform using the Function Generator.
80 | MTS Landmark Test System Operation
Running the Example HCF Test