a) Disable the force lower limit so that you are able to reposition the grip and reset the interlock.
b) Select Station Manager > Display > Station Setup > Channels > Ch1 (or other depending on

your configuration) > Force.

c) Click the Channel Input Signals button to display the Inputs: Axial Force window.
d) Set Lower Limit: Lower Action to Disabled.

3. Reset the Interlock.

In the Station Manager application main window, click the Reset button to clear the interlock.

4. Re-power the HPU and HSM.

In the Station Manager application main window, click the Power Low button, and then the Power
High
button for the HPU and HSM.

5. Reset force control to zero.

MTS Landmark Test System Operation | 103

Running the Example HCF Test