In the Station Manager application main window, click the Manual Control button. In the Manual
Controls window, reset Manual Cmd to 0.
6. Reset the meters.
Reset meter history by selecting Station Manager > Display > Meters and clicking the Reset button.
7. Reset the Lower (Compression) Limit.
a) Navigate to Station Manager > Display > Station Setup > Channels > Ch1 (or other depending
on your configuration) > Force > Limits.
b) Enter -0.05 kN in the Lower Limit text box and ensure that Interlock is selected for the Action.
8. Release the upper grip.
104 | MTS Landmark Test System Operation
Running the Example HCF Test