3 error codes: meaning and troubleshooting, Error codes: meaning and troubleshooting, Software – Leica Biosystems CM3600 XP User Manual
Page 76

76
Instructions for use, 6v3, RevB – 01/2013
6.8.3 Error codes: meaning and troubleshooting
Error code
Error message
Trouble shooting
Error 01
Error 02
Error 03
Error 04
Error 05
Error 06
Error 07
Error 08
Error 09
Error 10
Error 11
Knife holder – limit position
exceeded.
Limit position of sledge exceeded
Chamber – Alarm temperature
exceeded
Power failure
Data transfer failed (happens rather
frequently subsequent to a power
failure)
Emergency stop activated
Stepper motor – step error
Refrigerating system – pressure
exceeded
Flap limit switch malfunction
Call Technical Service
•
Re-initialize instrument. If the problem
recurs, call Technical Service.
•
Re-initialize instrument. If the problem
recurs, call Technical Service.
•
Check whether there has been a power
failure!
•
Check the selected alarm temperature
value. It must differ at least 5 K from the
chamber temperature selected.
•
Check the cable connection between PC
and control unit.
•
If problem recurs, call Technical Service
•
Not defined
•
Deactivate emergency switch
•
Call Technical Service.
•
Check the air circulation between the
condenser and the control unit.
•
Check the distance of the instrument to
the wall (at least 50 cm)!
•
If problem recurs, call Technical Service
•
Re-initialize instrument. If the problem
recurs, call Technical Service.
•
Call Technical Service.