5 refrigeration system - cooling chamber, Refrigeration system - cooling chamber, Characteristics of the leica cm3600 xp – Leica Biosystems CM3600 XP User Manual

Page 16

background image

16

Instructions for use, 6v3, RevB – 01/2013

200 V / 60 Hz, two-phase
Nominal voltage:

200 V AC

Nominal frequency:

60 Hz

Line voltage tolerance:

± 10 %

Power draw:

2400 VA

Automatic cutout/

2 x 8A / 1 x 16A

Power safety switch

UL-listed

220 V / 60 Hz, two-phase
Nominal voltage:

220 V AC

Nominal frequency:

60 Hz

Line voltage tolerance:

± 10 %

Power draw:

2400 VA

Automatic cutout/

2 x 8A / 1 x 16A

Power safety switch

UL-listed

2.2.5 Refrigeration system - cooling chamber

Selectable temperature range:

0° C to -30 °C

at an ambient temperature
of 22 °C, closed window
and humidity

≤ 60%:

-30 °C ±10 % (adjustable
in 1 K increments)

Control accuracy:

+1 to -3 K

Electrical power draw:

1160 W

Refrigerating output at -25 °C:

1370 W

Nominal pressure:

20 bar (maximum value)

Cutoff pressure:

25 bar (via Pressostat)

Safety factor:

2, as per UL 1262

Refrigerant:

1100 g, refrigerant R404A

Compressor oil:

1220 ccm alpha 22, Kyodo

Defrost:

Automatic hot gas defrost, programmable,
2 x automatic hot gas defrost / 24 hours.
Duration: 1 - 15 minutes, manual defrosting as needed.

2.

Characteristics of the Leica CM3600 XP