5 refrigeration system - cooling chamber, Refrigeration system - cooling chamber, Characteristics of the leica cm3600 xp – Leica Biosystems CM3600 XP User Manual
Page 16

16
Instructions for use, 6v3, RevB – 01/2013
200 V / 60 Hz, two-phase
Nominal voltage:
200 V AC
Nominal frequency:
60 Hz
Line voltage tolerance:
± 10 %
Power draw:
2400 VA
Automatic cutout/
2 x 8A / 1 x 16A
Power safety switch
UL-listed
220 V / 60 Hz, two-phase
Nominal voltage:
220 V AC
Nominal frequency:
60 Hz
Line voltage tolerance:
± 10 %
Power draw:
2400 VA
Automatic cutout/
2 x 8A / 1 x 16A
Power safety switch
UL-listed
2.2.5 Refrigeration system - cooling chamber
Selectable temperature range:
0° C to -30 °C
at an ambient temperature
of 22 °C, closed window
and humidity
≤ 60%:
-30 °C ±10 % (adjustable
in 1 K increments)
Control accuracy:
+1 to -3 K
Electrical power draw:
1160 W
Refrigerating output at -25 °C:
1370 W
Nominal pressure:
20 bar (maximum value)
Cutoff pressure:
25 bar (via Pressostat)
Safety factor:
2, as per UL 1262
Refrigerant:
1100 g, refrigerant R404A
Compressor oil:
1220 ccm alpha 22, Kyodo
Defrost:
Automatic hot gas defrost, programmable,
2 x automatic hot gas defrost / 24 hours.
Duration: 1 - 15 minutes, manual defrosting as needed.
2.
Characteristics of the Leica CM3600 XP