2 maintenance, 1 oiling, Oiling 110 – Leica Biosystems CM3600 XP User Manual
Page 110

110
Instructions for use, 6v3, RevB – 01/2013
8.2
Maintenance
1
2
3
8.
Cleaning and Maintenance
8.2.1 Oiling
Be sure to check the oil level every and before each section-
ing procedure.
Checking the oil level
Unscrew both lids (1). The oil level must always be above the small
capillary tubes (2).
Filling up the oil reservoirs
Unscrew both lids (1) and add low temperature oil PDP 40 (order no. -
see chapter 9.2). Screw both lids (1) back on.
Oiling the sledge guides
Once a month oil the sledge guides (3) with low temperature oil PDP 40
(order no. - see chapter 9.2).
The instrument requires maintenance from Leica Service at
an interval of once per year.
The pushbutton switch of the refrigeration must likewise be
pressed at least once a year.
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)