beautypg.com

JULABO FPW50-HL Refrigerated Circulators User Manual

Page 89

background image

HL

89

in commands:

Asking for parameters or temperature values to be displayed.

Command

Parameter

Response of circulator

version

None

Version number of the software (V X.xx)

status

none

Status message, error message (see page 91)

in_pv_00

none

Current bath temperature.

in_pv_01

none

Heating power being used (%).

in_pv_02

none

Temperature value registered by the external Pt100 sensor.

in_pv_03

none

Temperature value registered by the safety sensor.

in_pv_04

none

Setpoint temperature (“SafeTemp”) of the excess
temperature protection

in_sp_00

none

Working temperature “Setpoint 1”

in_sp_01

none

Working temperature “Setpoint 2”

in_sp_02

none

Working temperature “Setpoint 3”

in_sp_03

none

High temperature warning limit

“OverTemp”

in_sp_04

none

Low temperature warning limit

“SubTemp”

in_sp_05

none

Setpoint temperature of the external programmer
(REG+E-PROG).

in_sp_06

none

Temperature indication in °C or °F

in_sp_07

none

1. Adjusted pump stage in the -OFF- condition.
2. Pump stage corresponding to the effective rotational
speed after start.

in_sp_08

none

Value of a flow rate sensor connected to the E-Prog input

in_sp_09

none

Value of a pressure sensor connected to the E-Prog input

in_par_00

none

Temperature difference between working sensor and safety
sensor

in_par_01

none

Te - Time constant of the external bath.

in_par_02

none

Si - Internal slope

in_par_03

none

Ti - Time constant of the internal bath.

in_par_04

none

CoSpeed - Band limit (max. difference between the
temperatures in the internal bath and external system).

in_par_05

none

Factor pk/ph0:
Ratio of max. cooling capacity to max. heating capacity

in_par_06

none

Xp control parameter of the internal controller.

in_par_07

none

Tn control parameter of the internal controller.

in_par_08

none

Tv control parameter of the internal controller.

in_par_09

none

Xp control parameter of the cascade controller.

in_par_10

none

Proportional share of the cascade controller.

in_par_11

none

Tn control parameter of the cascade controller.