JULABO FPW50-HL Refrigerated Circulators User Manual
Page 12
Operator responsibility – Safety recommendations
12
FPW50-HL
Working temperature range
°C
-50 ... 200
Temperature stability
°C
±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20 -40
0.9 0.8 0.5 0.16
Refrigerant
R404A / R507
Overall dimensions
(WxDxH)
cm
42x49x72
Bath opening
(WxL)
cm
18x12
Bath depth
cm
20
Filling volume
liters
5.5 ... 8
Weight kg
57
Mains power connection
230 V/50 Hz
V/Hz
207-253 / 50
Current draw
(at 230 V)
A
14
Mains power connection
230 V/60 Hz
V/ Hz
207-253 / 60
Current draw
(at 230 V)
A
14
HL
Temperature selection
digital
via
keypad
indication on LCD DIALOG-DISPLAY (°C/°F)
remote control via PC
indication on monitor
Temperature indication
VFD COMFORT-DISPLAY (°C/°F)
Resolution
°C
0.01
ATC3
INT / EXT
°C
±3 / ±9
Temperature control
ICC - Intelligent Cascade Control
Heater wattage
(at 230 V)
kW
2.0
Heater wattage
(at 115 V)
kW
1.0
Electronically adj. pump capacity
stages
1 ... 4
Flow rate
at 0 bar
l/min
22 ... 26
Max. pressure
at 0 liters
bar
0.7
Max. suction
at 0 liters
bar
0.4
Electrical connections:
see page 13
Ambient temperature
°C
5 ... 40
Mains power connection 230 V/50-60 Hz
V/ Hz
190 ... 253 / 50-60
Current draw
(at 208 V / 230 V)
A
12 / 13
Mains power connection 115 V/50-60 Hz
V/ Hz
90 ... 127 / 50-60
Current draw
(at 100 V / 115 V)
A
9/10
All measurements have been carried out at: rated voltage and frequency
ambient temperature: 20 °C
Technical changes without prior notification reserved.