beautypg.com

JULABO FPW50-HL Refrigerated Circulators User Manual

Page 12

background image

Operator responsibility – Safety recommendations

12


FPW50-HL

Working temperature range

°C

-50 ... 200

Temperature stability

°C

±0.01

Cooling capacity
Medium: ethanol

°C
kW

+20 0 -20 -40
0.9 0.8 0.5 0.16

Refrigerant

R404A / R507

Overall dimensions

(WxDxH)

cm

42x49x72

Bath opening

(WxL)

cm

18x12

Bath depth

cm

20

Filling volume

liters

5.5 ... 8

Weight kg

57

Mains power connection

230 V/50 Hz

V/Hz

207-253 / 50

Current draw

(at 230 V)

A

14

Mains power connection

230 V/60 Hz

V/ Hz

207-253 / 60

Current draw

(at 230 V)

A

14



HL

Temperature selection

digital

via

keypad

indication on LCD DIALOG-DISPLAY (°C/°F)

remote control via PC

indication on monitor

Temperature indication

VFD COMFORT-DISPLAY (°C/°F)

Resolution

°C

0.01

ATC3

INT / EXT

°C

±3 / ±9

Temperature control

ICC - Intelligent Cascade Control

Heater wattage

(at 230 V)

kW

2.0

Heater wattage

(at 115 V)

kW

1.0

Electronically adj. pump capacity

stages

1 ... 4

Flow rate

at 0 bar

l/min

22 ... 26

Max. pressure

at 0 liters

bar

0.7

Max. suction

at 0 liters

bar

0.4

Electrical connections:

see page 13

Ambient temperature

°C

5 ... 40

Mains power connection 230 V/50-60 Hz

V/ Hz

190 ... 253 / 50-60

Current draw

(at 208 V / 230 V)

A

12 / 13

Mains power connection 115 V/50-60 Hz

V/ Hz

90 ... 127 / 50-60

Current draw

(at 100 V / 115 V)

A

9/10


All measurements have been carried out at: rated voltage and frequency

ambient temperature: 20 °C

Technical changes without prior notification reserved.