Technical specifications – JULABO FPW50-HL Refrigerated Circulators User Manual
Page 10

Operator responsibility – Safety recommendations
10
2.4. Technical
specifications
F25-HL
F32-HL
Working temperature range
°C
-28 ... 200
-35 ... 200
Temperature stability
°C
±0,01
±0,01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20
0.26 0.2 0.06
+20 0 -20 -30
0.45 0.39 0.15 0.06
Refrigerant
R134a
R134a
Overall dimensions
(WxDxH)
cm
23x42x64 31x42x64
Bath opening
(WxL)
cm
12x14
18x12
Bath depth
cm
14 15
Filling volume
liters
3 ... 4.5
5.5 ... 8
Weight kg
32 38
Mains power connection
230 V/50 Hz
V/ Hz
207-253 / 50
207-253 / 50
Current draw
(at 230 V)
A
12
12
Mains power connection
230 V/60 Hz
V/ Hz
-----
207-253 / 60
Current draw
(at 230 V)
A
-----
12
Mains power connection
115 V/60 Hz
V/ Hz
103-127 / 60
103-127 / 60
Current draw
(at 115 V)
A
13
14
Mains power connection
100 V/60 Hz
V/ Hz
90-110 / 50-60
90-110 / 50-60
Current draw
(at 100 V)
A
13
14
F33-HL
Working temperature range
°C
-30 ... 200
Temperature stability
°C
±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20 -30
0.5 0.32 0.12 0.03
Refrigerant
R134a
Overall dimensions
(WxDxH)
cm
36x46x71
Bath opening
(WxL)
cm
23x14
Bath depth
cm
20
Filling volume
liters
12 ... 16
Weight kg
45
Mains power connection
230 V/50 Hz
V/Hz
207-253 / 50
Current draw
(at 230 V)
A
12
Mains power connection
230 V/60 Hz
V/ Hz
-----
Current draw
(at 230 V)
A
-----
Mains power connection
115 V/60 Hz
V/ Hz
103-127 / 60
Current draw
(at 115 V)
A
15