5. Preparations

5.1. Installation



 Place the unit on an even surface on a pad made of non-flammable

material.

 Cooling machine, pump motor and electronics produce intrinsic heat

that is dissipated via the venting openings! Never cover these
openings!

 Keep at least 20 cm of open space on the front and rear venting grids.
 The place of installation should be large enough and provide sufficient

air ventilation to ensure the room does not warm up excessively
because of the heat the instrument rejects to the environment. (Max.
permissible ambient temperature: 40 °C).
For a fault (leakage) in the refrigeration system, the standard EN 378
prescribes a certain room space to be available for each kg of
refrigerant.
The refrigerant quantity is specified on the type plate.
> > For 0.25 kg of refrigerant R134a, a room space of 1 m

3

is required.

> For 0.52 kg of refrigerant R404A, a room space of 1 m

3

is required.

> For 0.49 kg of refrigerant R507, a room space of 1 m

3

is required.

 Do not set up the unit in the immediate vicinity of heat sources and do

not expose to sun light

 Before operating the unit after transport, wait about one hour after

setting it up. This will allow any oil that has accumulated laterally during
transport to flow back down thus ensuring maximum cooling
performance of the compressor.

 Set selector dial for cooling machine (23) in position "1" for operation

with HL circulator.

(only F25, factory setting)

Only water cooled models:

Ensure circulation of cooling water by connecting the tubing to cooling
water inlet and outlet on the rear of the refrigerated circulator.

 Cooling water connecter ¾"
 Cooling water see page 14



5.2. Bath fluids

Caution:

Carefully read the safety data sheet of the bath fluid used, particularly with regard to
the fire point!
If a bath fluid with a fire point of

65 °C is used, only supervised operation is possible.

Water:

The quality of water depends on local conditions.

 Due to the high concentration of lime, hard water is not suitable for temperature

control because it leads to calcification in the bath.