JULABO FPW50-HL Refrigerated Circulators User Manual
Page 11

HL
11
FP35-HL
FP40-HL
Working temperature range
°C
-35 ... 200
-40 ... 200
Temperature stability
°C
±0.01
±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20 -30
0.45 0.39 0.15 0.05
+20 0 -20 -40
0.68 0.5 0.32 0.04
Refrigerant
R134a
R134a
Overall dimensions
(WxDxH)
cm
31x42x66 36x46x71
Bath opening
(WxL)
cm
18x12
23x14
Bath depth
cm
----- 20
Filling volume
liters
2.5
9 ... 16
Weight kg
38 49
Mains power connection
230 V/50 Hz
V/Hz
207-253 / 50
207-253 / 50
Current draw
(at 230 V)
A
12
13
Mains power connection
230 V/60 Hz
V/ Hz
-----
207-253 / 60
Current draw
(at 230 V)
A
-----
14
Mains power connection
115 V/60 Hz
V/ Hz
103-127 / 60
-----
Current draw
(at 115 V)
A
15
-----
Mains power connection
100 V/60 Hz
V/ Hz
90-110V/50-60Hz
-----
Current draw
(at 100 V)
A
14
-----
FP45-HL
FP50-HL
Working temperature range
°C
-42 ... 200
-50 ... 200
Temperature stability
°C
±0.01
±0.01
Cooling capacity
Medium: ethanol
°C
kW
+20 0 -20 -40
0.85 0.7 0.42 0.08
+20 0 -20 -40
0.9 0.8 0.5 0.16
Refrigerant
R-404A
R404A / R507
Overall dimensions
(WxDxH)
cm
38x58x69
42x49x72
Bath opening
(WxL)
cm
23x26
18x12
Bath depth
cm
20
20
Filling volume
liters
18 ... 26
5.5 ... 8
Weight kg
53
57
Mains power connection
230 V/50 Hz
V/Hz
207-253/50-60
207-253 / 50
Current draw
(at 230 V)
A
13
14
Mains power connection
230 V/60 Hz
V/ Hz
207-253/50-60
207-253 / 60
Current draw
(at 230 V)
A
13
14