Safety installations, warning functions

34

8.0.1.

Early warning system, low level protection

DBGM: 20306059.8

This low level protection is independent of the control circuit and is
divided into two sections:

1. Switch in stage 1 recognizes a critical fluid level

.

An audible warning (interval tone) sounds and a message
appears on the VFD COMFORT-DISPLAY.

Refill bath fluid!

2. Switch in stage 2 recognizes a low fluid level

.

If stage 2 of the low level protection device (according to IEC
61010-2-010) is triggered, a complete shutdown of the heater
and circulating pump is effected.
A continuous alarm tone sounds and a message >ALARM<
>CODE 01< appears on the VFD COMFORT-DISPLAY.

Press

for help on the LCD 


Turn off the unit with the mains switch, refill bath fluid and turn
the unit on again!

3. Float

4. Circulating

pump

5. Heater

Warning:

For refill always use the same bath fluid type that is already in the bath.
Bath oils must not contain any water contaminants and should be pre-heated to the
actual bath temperature! Explosion hazard at higher temperatures!