beautypg.com

JULABO FPW50-HL Refrigerated Circulators User Manual

Page 75

background image

HL

75

9.9.2.

Outputs of the connector - REG+E-PROG


Outputs of the connector

1. First define the desired output value for channels 1 to 3:


ActInt

internal actual temperature value (bath temperature)

ActExt

external actual temperature value (external sensor)

Power

periodic or intermittent heating or cooling

Setpoint

active setpoint temperature

(setpoint 1, 2, 3/integr. programmer/external programmer)

2. Then select the scale for channels 1 to 3:


Voltage outputs channels 1 and 2

Assign the voltage values of 0 V to the lowest and 10 V to the highest
necessary temperature or power rating required as an output value
(°C/%).
Current output channel 3

Assign the current values 0 mA or 4 mA to the lowest and 20 mA to
the highest temperature or power rating required as an output value
(°C/%).

3. The current output (channel 3) offers 2 ranges for selection:

0 to 20 mA or 4 to 20 mA

0 V

0 mA

20 mA

4 mA

10 °C

210 °C

197 °C

202 °C

 5 K

 200 K

Examples:

lowest temperature value:

10 °C

highest temperature value 210 °C

Fig.

shows

200

C scaled to paper width

slope: 50 mV/°C


lowest temperature value:

197 °C

highest temperature value: 202 °C

Fig.

shows

5

C scaled to paper width

slope:

2000

mV/°C