Page 43
HL
43
Editing
PrgNr: Set the program you wish to create. 0 ... 5
SP :
Setpoint
TI :
Time [hh:mm] or
GRD: Gradient [°/t] = [°/min]
Change type of program in MENU >configuration<,
menu item >Programmtyp/program type< .
ST:
Step 0 … 59
Press the
key if a parameter is to be retained. (prior to OK)
Level 2
Example:
Program (PrgNr.) 2 is to have a target temperature with a set
point (SW) of 100.00 °C in section t (ABS) 3.
Duration (ZI) 1 hour.
Select the program number.
Set the desired >PrgNr.< by pressing
and
.
Level 3
For the selected section >ST< a setpoint >SP< and the
duration >TI< or the gradient >GRD< are entered in the
right half of the input window.
or
Input window for >ST<, >SP< and >TI< / >GRD<
1.
Set the desired Step >ST < by pressing
and
.
2.
Set set point >SP< by using the numeric keypad and
confirm by pressing
.
3.
Set the time >TI< or a gradient >GRD< using the numeric
keypad and confirm by pressing
.
By confirming time or GRD the values on the left side of
the input window are moved.
The values for the next section are entered in the right half
of the input window.
See above 1. / 2. / 3.