JULABO FPW50-HL Refrigerated Circulators User Manual

Page 66

background image

Menu functions

66

The value >TmpValue< is only indicated.

In addition the measured temperature
value >CalValue X< is saved during the
next step.

 The value flashes. Set by using the

numeric keypad and confirm by pressing

.


Return to 2. Set working temperature value: 120.00 °C

The value >TmpValue< is only indicated.

In addition the measured temperature
value >CalValue X< is saved during the
next step.

 The value flashes. Set by using the

numeric keypad and confirm by pressing

.

Return to 2. Set working temperature value: 160.00 °C

The value >TmpValue< is only indicated.

In addition the measured temperature
value >CalValue X< is saved during the
next step.

 The value flashes. Set by using the

numeric keypad and confirm by pressing

.

5. Return to standard operation

Set >YES< after calibration.
(Standard operation)

 The parameter flashes, switch by

pressing

and