9.9.5. Alarm-output

ALARM


Alarm output (10)
(for external alarm signal)
This socket is a potential-free change-over contact. With the
adjustments in the menu >Inputs/Outputs<, all operating conditions
can be signaled without having to change the pin assignments.

Meaning of the terms under menu item >Function< :

The circulator is in condition

>StandBy< or

>Alarm<

For >Type: normal<, pins 2 and 3 are connected in any case
according to the selected >Function<.

For >Type: inverse<, pins 2 and 1 are connected in any case
according to the selected >Function<.

Switching capacity

max. 30 W/40 VA

Switching voltage

max. 125 V

/

Switching current

max. 1 A