9.5.2. Selftuning

When performing a selftuning for the controlled system (temperature
application system), the control parameters Xp, Tn and Tv are
automatically determined and stored.

Possible parameters:

off

- no selftuning

The control parameters ascertained during the last identification are
used for control purposes.

once

- single selftuning (factory setting)

The instrument performs a single selftuning of the controlled system
after each start with the

key or after receiving a start

command via the interface.

always

- continual selftuning

The instrument performs a selftuning of the controlled system
whenever a new setpoint is to be reached.
Use this setting only when the temperature application system
changes permanently.

9.5.3. Dynamik

intern

This parameter affects the temperature sequence only in case of internal
control.

°C

t

aperiodic

normal

setpoint

temp. stability

°C

t

aperiodic

normal

temperature ramp

temp. stability

Factory setting:

APER (aperiodic)


Possible parameters:

normal

Allows for reaching the setpoint faster – with

setpoint change or ramp function – but
overshooting of up to 5 % is possible.

aperio

Ramp function: the increase of temperature

occurs temporally offset and achieves the target
temperature without overshooting.
Setpoint change: The temperature increases at
the same rate, the target temperature is achieved
without overshooting.

With both settings constant temperature is achieved
after approximately the same time.