Restore parameter store (instruction 20h), Remotely activate (instruction 30h), Remote monitoring – INFICON Maxtek PM-700 Plating Monitor User Manual
Page 62: Able, Emote, Ctivation, Ommand, Odes

PM-700 SERIES PLATING MONITOR
54
4.8.4.3.2
Restore Parameter Store (Instruction 20h)
This operation inputs the entire parameter store from the host computer to the
plating monitor and then stores these values in their respective memory locations.
To initiate this operation, the computer would output the following data stream:
Header
(2 bytes)
Instruction Code
(1 byte)
Message
Length
(1 byte)
Command
Code or
Message
Checksum
(1 byte)
FFh FEh
20h
47h
71
bytes
checksum
4.8.4.3.3
Remotely Activate (Instruction 30h)
This operation allows the computer to perform a remote key press of the plating
monitor's front panel buttons through the RS232. The valid key codes are as
follows:
Table 9 Remote Activation Command Codes
Remote Activation
Command Code
Description
1h
Set DAC outputs to zero scale
2h
Set DAC outputs to full scale
4h
Manual probe swap *
8h Stop
button
10h Deplate
button
20h Preplate
button
40h Plate
button
* This parameter is only available in the dual probe models.
The data stream format to initiate a remote activation command is as follows:
Header
(2 bytes)
Instruction Code
(1 byte)
Message
Length
(1 byte)
Command Code
or Message
Checksum
(1 byte)
FFh FEh
30h
1h 1
byte
checksum
For example, to initiate a remote plate start, the computer would send: