INFICON CrystalSix Sensor User Manual
Page 44

4 - 8
IP
N 07
4-
15
5L
CrystalSix Operating Manual
5
Place the compression spring (item #28) in the cavity surrounding the 0.25"
diameter shaft of the carousel assembly. Install the Teflon washer (item #39)
onto the.25" diameter shaft.
6
With the connector facing outward, pass the top plate (item #10) over the 0.25"
diameter shaft of the carousel assembly (item #5) and the .125" diameter
alignment pin.
7
Secure the top plate to the sensor body utilizing the three #4-40 screws (items
#14 and 17).
8
Install the pawl and actuator stem assembly (item #24) into the bellows
assembly (item #19) by threading the actuator stem into the mating hole
located in the bore of the bellows assembly. Continue to thread the actuator
stem until the shoulder of the actuator stem is approximately .035" away from
the bellows assembly. This .035" distance specification is for reference only;
the actuator stem position may have to be adjusted to obtain the desired
condition, as illustrated in
9
Fasten the bellows assembly to the top plate utilizing the two #4-40 x 1.125"
long screws (item #20). Do not tighten the screws. Position the bellows
assembly in such a fashion as to allow the actuator stem to come in contact with
the ball bearings (item #25) as illustrated in
. Tighten the
screws. Loosen the #2 screw (item #42) which secures the detent to the bellow
assembly (item #19).
10
Slide the ratchet (item #22) and the stop ratchet (item #38) onto the 0.25"
diameter shaft of the carousel assembly (item #5) as illustrated. Do not tighten
the #6-32 set screws (item #33).
11
Attach the loop of the extension spring in the hole provided in the pawl of the
pawl and actuator stem assembly (item #24).
12
Position the ratchet (item #22) such that it engages the pin of the pawl and
actuator stem assembly (item #24).
13
Position the detent (item #41) so it engages the ratchet (item #22).
Unit is now ready to be aligned, refer to
.