CAUTION
Incorrect rotation direction with the stops disabled will
cause permanent damage rendering transducer
inoperative. Refer to
for proper
direction of rotation.
The unit, when properly aligned, will index the crystal to a position that is centered
within the aperture provided in the heat shield. The unit must be aligned if this
condition does not exist.
Reference
.
Equipment Required:
5/64" allen wrench (Included in 750-268 Kit)
1/16" allen wrench (Included in 750-268 Kit)
3/32" allen wrench (Included in 750-268 Kit)
Alignment tool assembly 750-254-G1 (Included in 750-268 kit)
Regulated air supply 80-90 PSIG (5.5 bar - 6.2 bar) [550 kPa - 620 kPa]
NOTE: 75 PSIG (5.2 bar) [520 kPa] for alignment purposes only)
4.4.1 Alignment Instructions
With the required equipment at hand, proceed as follows:
1
Temporarily connect an air supply to the bellows assembly supply tube.
Regulate the air pressure to 75 PSIG (5.2 bar) [520 kPa].
2
Remove the actuator cover (item 15) and related hardware.
3
Insert the 0.125" diameter alignment pin (of the 750-254-G1 assembly) in the
alignment hole (if not already installed) as shown in
The carousel may need to be rotated to achieve the alignment condition shown
in
4
Loosen the set screws (two set screws per item) of the ratchet (item #22) and
the stop ratchet (item #38).
5
Loosen the #2 fastener (item #42) that attaches the detent (item #41) to the
bellows assembly item #19). Do NOT remove detent.
6
Actuate the bellows assembly by applying 24 volts to the leads of the solenoid
valve of the solenoid valve assembly. Consult
for wiring
assignments. The air pressure must be sustained.