CrystalSix Operating Manual

3.2 Installation with 2 3/4” Feedthrough

Installation of the solenoid valve requires a 2 3/4" feedthrough inclusive of two
coaxial feedthroughs (IPN 002-080, see

Figure 5-2 on page 5-2

). The second

coaxial feedthrough is not used, and should be protected from damage as a result
of process material. Follow the steps below:

1

Install the Feedthrough.

2

Add the valve bracket (modified) to the desired location as illustrated in

Figure

3-3 on page 3-6

.

3

Tighten the feedthrough bolts.

4

Install the air fitting to the female thread adapter.

5

Connect 1/8" diameter tubing from the valve output (A) to the feedthrough
fitting. See the CAUTION in

section 3.1

above.

6

Attach the valve’s supply (P) (normally closed (NC) port) to the 80-90 PSIG
(5.5 bar - 6.2 bar) [550 kPa - 620 kPa] source of air. Verify that the orifice
(IPN 059-189) has been installed into this NC valve port. (See Figure 3-1b of

Figure 3-1 on page 3-4

.)

NOTE: Maximum temperature for the solenoid valve assembly is 105 °C for

bakeout and operation.

3.3 Electrical and Pneumatic Connections

3.3.1 Electrical

To complete installation of the assembly, make electrical connections where
indicated in

Figure 3-3 on page 3-6

to either 24 V(ac) or V(dc). Current required is

approximately 70 mA.

CAUTION

Maximum applied voltage must not exceed 26 V(ac).