3.2 Installation with 2 3/4” Feedthrough
Installation of the solenoid valve requires a 2 3/4" feedthrough inclusive of two
coaxial feedthroughs (IPN 002-080, see
). The second
coaxial feedthrough is not used, and should be protected from damage as a result
of process material. Follow the steps below:
1
Install the Feedthrough.
2
Add the valve bracket (modified) to the desired location as illustrated in
3
Tighten the feedthrough bolts.
4
Install the air fitting to the female thread adapter.
5
Connect 1/8" diameter tubing from the valve output (A) to the feedthrough
fitting. See the CAUTION in
6
Attach the valve’s supply (P) (normally closed (NC) port) to the 80-90 PSIG
(5.5 bar - 6.2 bar) [550 kPa - 620 kPa] source of air. Verify that the orifice
(IPN 059-189) has been installed into this NC valve port. (See Figure 3-1b of
NOTE: Maximum temperature for the solenoid valve assembly is 105 °C for
bakeout and operation.
3.3 Electrical and Pneumatic Connections
3.3.1 Electrical
To complete installation of the assembly, make electrical connections where
indicated in
to either 24 V(ac) or V(dc). Current required is
approximately 70 mA.
CAUTION
Maximum applied voltage must not exceed 26 V(ac).