Technical Data
79
ninp69e1-a (1410)
11.4.6
Control Output
Pin
Signal
Type
Load
Comment
1
+24 VDC
SUPPLY
300 mA
Tooling switch supply (e.g. proximity
switch).
2
TS1
IN
-7 mA
Tooling switch 1.
3
TS2
IN
-7 mA
Tooling switch 2.
4
TS3
IN
-7 mA
Tooling switch 3.
5
TS4
IN
-7 mA
Tooling switch 4.
6
MARKER*
OUT
0.5 A
Marker output. Selectable mark on
REJECT or ACCEPT.
7
COM
GND
-1.0 A
Common GND.
8
COM
GND
-1.0 A
Common GND.
*
MARKER output (Tooling Connector, pin 6) can be used to send a start pulse to
marking equipment such as an engraving machine or a valve controlling a simple
pneumatic stamp. Function and length of pulse is set by the following two
parameters:
Marker Output:
Length of marker output pulse.
Output will go high at end of gas test and stay
high for the given time.
Marker Output High if Leak:
Decides function of marker pulse.
To mark rejected part set to OFF.
To mark accepted part, set to ON.
Connector:
8 pin male Weidmüller, Omnimate BL3.5. Mating screw terminal
included.
Purpose:
External start and stop. Control of optional external valves.
Pin
Signal
Type
Load
Comment
1
+24 VDC
SUPPLY
2.0 A
Start and stop switch and supply.
2
EXTSTART
IN
-7 mA
Start button return (NO contacts) or
contact to +24 VDC.
3
EXTSTOP
IN
-7 mA
Stop button return side (NO contact) or
contact to +24 VDC.