Troubleshooting
54
ninp69e1-a (1410)
Tracer Gas Fill Valve
The step tests the gas fill valve for internal leakage by registering pressure rise behind
the valve.
Possible results:
•
No Internal Leakage
•
Internal Leakage
Internal leakage can result in:
•
Erroneous pressure decay results
•
False vacuum decay rejects increased gas consumption
Test Port 2 Valve
This step tests Test Port 2 valve for internal leakage by registering pressure rise behind
the valve.
Possible results:
•
No Internal Leakage
•
Internal Leakage
Internal leakage can result in:
•
False blockage test accepts
Hardware Test
CONTINUE
STOP
Vacuum:
Pressure:
0.00 bar
bar
Tracer Gas Fill Valve
No Internal Leakage
0.00
Hardware Test
CONTINUE
STOP
Vacuum:
Pressure:
0.00 bar
bar
Test Port 2 Valve
Valve works!
0.00
Notice
This test will fail if both test ports are connected to a test object. Proceed
and then repeat the entire hardware test sequence with both ports
plugged to perform this test step.