Troubleshooting

54

ninp69e1-a (1410)

Tracer Gas Fill Valve

The step tests the gas fill valve for internal leakage by registering pressure rise behind
the valve.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

Erroneous pressure decay results

False vacuum decay rejects increased gas consumption

Test Port 2 Valve

This step tests Test Port 2 valve for internal leakage by registering pressure rise behind
the valve.

Possible results:

No Internal Leakage

Internal Leakage

Internal leakage can result in:

False blockage test accepts

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Tracer Gas Fill Valve

No Internal Leakage

0.00

Hardware Test

CONTINUE

STOP

Vacuum:

Pressure:

0.00 bar

bar

Test Port 2 Valve

Valve works!

0.00

Notice

This test will fail if both test ports are connected to a test object. Proceed
and then repeat the entire hardware test sequence with both ports
plugged to perform this test step.