INFICON Sensistor ILS500F Leak Detection System User Manual
Page 61
61
Troubleshooting
ninp69e1-a (1410)
Tested Unit
Tested Feature
Reason for Error
Action
Tracer Gas Fill Valve
Internal leaks
Dirty or worn valve seals.
Replace or clean tracer
gas fill valve.
Leaking pilot valve.
Replace fourth valve from
bottom in pilot valve ramp.
Test Port 2 Valve
Function
Dirty or broken pilot valve. Replace third valve from
bottom in pilot ramp.
Test port 2 valve broken.
Replace valve.
Tracer Gas Fill Valve
Function
Dirty or broken pilot valve. Replace fourth valve from
bottom in pilot ramp.
Tracer gas fill valve
broken.
Replace valve.
Gas Valve Manifold
Leaks to outside
Wrongly assembled gas
valve.
Remove leaking valve.
Clean and grease valve
seal before installing
again. See instructions.
Wrongly installed
connectors/plugs.
Remove leaking unit.
Clean and grease o-ring
Install again.
Units lacking o-ring seal
should be sealed with
Loctite 577 or similar.
Evacuation Valve
Function
Dirty or broken Evacuation
pilot valves.
Replace fourth valve from
bottom in pilot valve ramp.
Lamp
Function
Broken lamp.
Replace lamp.
Send in for repair.
Tooling Valves
Function
Dirty or broken pilot valve. Replace first or second
valve from bottom in pilot
ramp.
Button
Function
Broken switch.
Send in for repair.